Available delivery methods: Home Parcel, Delivery agreed 9-21, Postnord Daytime delivery to door 8-16, Pick up from a Postal outlet, Delivery to Door, Delivery to Door with DPD Classic, Delivery to Door with DPD Classic, Delivery to Door with DPD Classic, Delivery to Door with DPD Classic, Delivery to Door with DPD Classic, Delivery to Door by Parcelforce, Delivery to Door with Postnord Parcel, Delivery to Door with Postal EMS, Delivery to Door with Postal EMS service, Pick up from a MyPack Collect point, Delivery to MyPack Collect Service Point, Toimitus ja asennus asiakkaan tiloihin, PKS-alue, Pick-Up from Vantaa office, Nouto - ei veloitusta

CALIBRATION OF THE pH & EC METER
 

EC and pH measurements should be calibrated every few months. There is included screw driver for this with the meter.

 

Calibrate the meter using pH buffer agents and EC calibrators with this Kit. 

 

The Kit includes ready-made calibration liquids pH7, pH4 and EC 2.8. In addition, the package includes pH calibration powders, with which you can prepare more pH calibration liquid by adding 250 ml of pure water to the powder, e.g. battery water, which can be obtained inexpensively from car dealerships.

 

Calibration of the meter to pH 7

  • Open the screw cap on the meter, which shows the batteries

  • Use the EC / pH button to select the pH to be measured, the display will show a low PH

  • Remove the probe cap, immerse the probe in the pH 7 solution

  • Wait for the meter to stabilize, then adjust the meter to 7.00 by turning the small PH screw next to the batteries.

  • Rinse the meter head with water and replace the protective cap

 
Calibration of the meter to pH 4

  • By also calibrating the meter to pH 4, you ensure that the meter is calibrated over the entire measuring range

  • Remove the probe cap, immerse the probe in the pH 4 solution

  • Wait for the meter to stabilize, then adjust the meter to 4.00 by turning the small PH screw next to the batteries.

  • Rinse the meter head with water and replace the protective cap

 

 

EC calibration of the meter

  • Open the screw cap on the meter, which shows the batteries

  • Select EC to measure with the EC / pH button, the display will read X10 US with a small

  • Remove the probe cap, immerse the probe in the EC / CF calibration solution

  • Wait for the meter reading to stabilize, then adjust the meter to 280 by turning the small TDC / EC screw next to the batteries.

  • Rinse the meter head with water and replace the protective cap

This includes the following individual products

VitaLink CF Standard 250ml - ESSENTIALS Calibration Fluid - 2.8mS for adjusting EC meter with Hydroponics growing

1 piece(s)
6.90
Delivery weight: 0.3 kg

pH 4 calibration buffer with for calibrating pH meter to pH 4.0 with hydroponics |aquarium | swimming pool | well water

1 piece(s)
6.90
Delivery weight: 0.3 kg

pH calibration buffer with pH 7

1 piece(s)
5.90
Delivery weight: 0.3 kg

Powder to calibrate the pH meter. Make a pH 4.01 calibration solution | pH calibration fluid with 250 ml water.

1 piece(s)
2.00
Delivery weight: 0.01 kg

Powder to calibrate the pH meter. Make a pH 6.86 calibration solution |pH 7 calibration fluid with 250 ml water.

1 piece(s)
2.00